CQRP Club LC Meter — Alignment and Usage

CQrp Club Project - PIC L/C Meter

Alignment and Usage

CQRP Club LC Meter — Alignment and Usage — 5 rows, as published. Scroll sideways on a narrow screen.
Click on the image for a larger view. There are only two adjustments which need to made to the meter board, setting the contrast for the LCD display and stabilizing the inductance oscillator. Both of these adjustments are simple and require no test equipment. To adjust the LCD contrast setting, turn the 10K pot fully counter-clockwise and turn the meter on. Looking at the LCD display, you should see dark squares across at least the top line. Turn the pot clockwise until these squares fade in to the background and adjust the contrast to make the displayed characters most readable to suit you. Remember that the viewing angle is important when making this adjustment.
Click on the image for a larger view. The second adjustment is made to stabilize the oscillator in the inductance measuring circuit. In the original article, the author indicated that this adjustment is not critical but I have found that not to be the case. This may be caused by the additional 68pF capacitor that I have installed in the circuit, but in any case, it does have a definite effect on the meter operation. To adjust the inductance oscillator circuit, turn the power off and set to 10K pot adjustment fully clockwise. Short the two alligator clip leads together and turn on the meter. Make sure that you are not touching the clip leads while making any adjustments. On the top line of the LCD display you will see the frequency that the oscillator is runing at and on the second line will be a number displayed in uH. Adjust the pot slowly counter-clockwise to obtain a reading around 12uH (if you swing the pot from one stop to the other, you should get readings between 10 and 18uH.). Notice the frequency being displayed in the first line. It should be between 350KHz and 500KHz and it should be bouncing up or down in fairly large steps. Turning the pot clockwise, in very small increments, the frequency should start to stabilize and you should be able to find a point where it changes by less than 50Hz up or down. When you have found this point, zero your meter by pressing the momentary switch. The top line will briefly display NULLED and then the display will show the frequency on the top line and the numbers 0.000uH on the second line. Unshort your clip leads for a few seconds and then clip them back together. The readout should once again show a value of 0.000uH when it settles. Do this 3 or 4 times and if the value returns each time to 0.000uH, you have successfully adjusted your meter. If the zeroed value changes, then you need to readjust the pot until there is no change when making the above tests. Don't be concerned about the displayed frequency itself, it may be anywhere from 300kHz to 500KHz, and no two meters will display the same frequency. This is due to lead lengths, internal wiring variations, etc. within each individual meter. However, the displayed inductance value, before zeroing should be between 10 and 12uH. This completes all internal adjustments to your meter. All further adjustments are done in the software within the PIC.
To calibrate your meter you are going to require at least a couple of inductors and capacitors of known value. It is preferable to have several test components of both large and small values. Obviously the lower the tolerance value of the calibration sources (1% or 2% vs 5% or 10%), the closer you will be able to calibrate your meter. The software in the PIC has the ability to allow you to compensate for component variances, lead lengths, etc within your meter. To calbrate the Inductance meter, turn your meter off and then turn it back on. Short the two clip leads together and zero, or null, your meter by pressing the momentary switch. Unshort the clip leads and measure one of your test coils of known value. Note the inductance reading. The actual value at this point is not important. Remove the first coil, short your leads and insure that your meter is still zeroed. Measure a second coil and again note the value. You can repeat this with as many test inductors as you like. The values displayed will either be higher or lower than the actual known value. To calibrate your meter, turn the meter off. Holding the momentary switch down, turn the meter back on to the inductance position. After initialization, the display should show IND CORRECTION and display a value of 100 on the second line. Release the momentary switch. Depending on whether your measured values were higher or lower than the known value, you will need to increment or decrement this value. In other words, if your test inductor was 10uH and your measured value was 9.1uH, you are going to want to increase the offset number. According to the author, each step is 1%, with the range being from 1 to 199. To increment (increase) the offset value, move the rotary switch to the CAP position and press the momentary switch. The number displayed will increase. To decrement (decrease) the offset value, turn the rotary switch to the IND position and press the momentary switch. The number will decrease. For large jumps it is not necessary to press the momentary switch for each increase/decrease after the first press. Holding the momentary switch closed will cause the number to increase or decrease until released or reaching 199 or 1. Turning off your meter saves this value to the PIC and it will remain there until changed again, regardless of whether power is applied to the meter or not. Turn you meter back on to the inductance position, zero your meter and again measure your test coils. Repeat this process until the displayed value is as close to the known value as you can get.
Calibrating your Capacitance meter is the same process as above. The one difference is that you do not short the clip leads together when zeroing your meter - leave them open. After zeroing your meter, measure a capacitor of known value and note the reading. Do this again with a second and third capacitor of different values. To adjust the capacitance offset, turn your meter off. Holding the momentary switch closed, turn your meter back on the Capacitance position. When CAP CORRECTION appears in the display, release the momentary switch and proceed to set your offset value as required. Again, with the switch in the CAP position, pressing the momentary switch will increase the value. Moving the rotary switch to the IND position and pressing the momentary switch will decrease the offset value.
Using your meter is simple. Just remember that each time you are going to measure either a coil or capacitor, you must zero your meter. In measuring inductors, turn the rotary switch to the IND position, short the clip leads together, wait a few seconds for the inductance value displayed to settle and press the momentary switch. Then measure your coil, keeping your hands away from the leads and component being measured. To measure a capacitor, turn the rotary switch the CAP position and insure that the clip leads are NOT shorted. Zero your meter and measure your cap, again keeping your hands away from the leads and component.

Close up your enclosure and stick a fork in yourself - you're DONE!!!